Hoyer Power Patient Lifters P-C-HLA-2 & HPL402 User Instruction Manual

Hoyer Power Patient Lifters P-C-HLA-2 & HPL402 User Instruction Manual

ASSEMBLY
The following tools are required for assembly: HPL402: 3/4″ & 1/2″ wrench, 3/16″ hex wrench, 1/8″ hex wrench, and a thin layer of grease should be applied to all bolts. P-C-HLA-2: two 9/16″ wrenches, 1/8″ hex wrench, and a thin layer of grease should be applied to all bolts.

1. Carefully unpack Hoyer Lifter. Check for 6-point cradle and adjusting lever handle.
2a.P-C-HLA-2– Insert mast in base sleeve (Fig. 1). Be sure notch on end of mast fits over bolt in bottom of mast sleeve. Mast must lock into position in base sleeve, making it impossible to rotate. Check carefully to be certain mast cannot rotate.
2b. HPL402– Insert mast in base and tighten bolt.
3. Make certain boom points in same direction as base legs. Thread mast safety lock and stud into threaded hole in mast sleeve. Tighten firmly, locking mast into base (Fig. 2). Keep mast and base locked at all times except when removing mast from base for storage. HPL402 only– all shoulder bolts must engage through bracket on both sides. If bolt is not fully engaged, misalignment and premature failure may result. Worn parts could fail, resulting in injury to resident/caregiver.
4. a. P-C-HLA-2 only– Insert boom end into yoke hinge at the top of mast. Fasten bolt and nut (Fig. 3). Be certain hex of bolt head fits into formed hex on side of hinge and that plastic washer is under lock nut.
b. HPL402– Attach boom to top of mast with hardware (Fig. 4)

Get pdf Hoyer Power Patient Lifters P-C-HLA-2 & HPL402 User Instruction Manual